The female testosterone range on your printout is the product of one method and one group of tested women, and its lower edge is set by the machine.

The case file

  1. Reassessment of the Access Testosterone chemiluminescence assay and comparison with LC-MS method.J Clin Lab Anal authors, Beckman DxI platform · 2018The female interval was below 0.5 to 2.78 nmol/L, derived indirectly from 2,838 outpatients; the limit of quantitation was 0.53 nmol/L.
  2. The Simultaneous measurement of serum testosterone and 5α-dihydrotestosterone by gas chromatography-mass spectrometry (GC-MS).Single-laboratory GC-MS method, Clin Chim Acta · 2018"a bias in favour of immunoassays at low concentrations for T in females"
  3. Multicenter evaluation of a new automated electrochemiluminescence immunoassay for the quantification of testosterone compared to liquid chromatography tandem mass spectrometry.Multicentre evaluation, Clin Biochem · 2011"Improved agreement with LC-MS/MS was demonstrated, especially for female patients."
  4. Salivary testosterone reference intervals for adult women and men.Clin Chem Lab Med study, LC-MS/MS measurement · 2026In 223 adult women, morning saliva range was 5.5 to 49.0 pmol/L and evening below 27.5 pmol/L; saliva values, not serum.

Audit your printout in five minutes

A printout or screenshot of your result, an older one if you have it, and a pen (5 min) (5 min) · 5 min

  1. Find the method or platform name (an immunoassay, mass spectrometry, or a machine name). Write it down, or write "not stated".
  2. Write the units: nmol/L or ng/dL. To convert nmol/L to ng/dL, multiply by 28.8.
  3. Write the lowest measurable value, often printed like "<0.5". If your result carries a < sign, write it as "below what the method sees", not zero.
  4. Write where the range comes from: the laboratory's own patients, the manufacturer's leaflet, or not stated.
  5. Count how many of the four you found, 0 to 4. With an older result, check that method and units match. If they do not, the two numbers cannot be compared.

What you end up withlines found on the printout, out of four

  • 4 of 4clearThe printout is readable: you know the method, the units, the floor and where the range comes from.Ask one question: was the range built with this same method on this laboratory's own patients?
  • 3 of 4mildOne line is missing, so part of the number's passport is blank.Phone the laboratory and ask for the missing line.
  • 1 or 2 of 4moderateMost of the passport is missing. The number cannot be matched to a range with confidence.Ask in writing for the method and the source of the range.
  • 0 of 4markedA number with no passport. Nothing on the page says which range it belongs to.Request the full report, not the portal summary.

When this test liesThe test misleads on a portal summary with the footnotes cut off, when the laboratory changed platform between your two results, when you compare total with free testosterone, and with saliva figures, which are a different sample [[41996543]]. It does not tell you whether the result is normal.

Find out who the women behind the range were

The best-documented female range I could trace was built like this. A laboratory on the Beckman DxI platform took routine results from 2,838 female outpatients and 2,445 male outpatients, and kept the middle 95 of every 100 results 1. For women that band ran from below 0.5 to 2.78 nmol/L, nanomoles per litre, the unit of concentration 1.

Nobody recruited healthy women. The range was worked out indirectly, from results that already existed because a clinician had ordered the test 1. Every woman in that sample had a reason to be sampled.

Cut that way, five results in every hundred land outside the band by construction.

"Normal" on your printout means "this is what turned up in women who were sent for a blood test".

Read the lower edge as the limit of the machine

Look at the left end of that range: below 0.5. The same platform measures reliably only from 0.53 nmol/L upward, and even there a result can sit about a fifth away from the true value 1.

So "below 0.5" says the instrument cannot see the number. It does not say you have little. A mass-spectrometry method in another laboratory reported a measurable floor of 0.117 nmol/L 2. The floor belongs to the method.

If your units are ng/dL, multiply nmol/L by 28.8 to convert. That changes the units. It does not make two methods agree.

See which way each machine errs, and why you cannot guess

An immunoassay uses antibodies that grab the hormone. Mass spectrometry weighs the molecules and is the stricter reference. Women's levels are low, and low levels are where antibodies struggle.

Two papers report opposite directions. In one laboratory's comparison with gas chromatography mass spectrometry, immunoassays read higher at low female concentrations 2. In the Beckman reassessment, the immunoassay read slightly lower than liquid chromatography mass spectrometry, at about 0.906 of the reference value in serum 1. The platforms and the concentration ranges differ, so no rule of thumb fits both.

Generation matters too. A 2011 multicentre evaluation found that a newer immunoassay agreed better with mass spectrometry than the former one did, and the authors called the gain especially clear in women 3. If your laboratory changed platform between two tests, the change may be the machine.

Authors who built a mass-spectrometry method in 2008 gave their reason: commercial direct immunoassays are sometimes inaccurate in women and children 4.

The Beckman comparison rests on 40 samples 1, so do not bank on its direction alone. I found no study giving a female serum range from a directly sampled, representative group measured by mass spectrometry. Until one exists, no printed range is the truth.

Do not borrow a range from another kind of sample

A 2026 study measured saliva in 223 adult women by mass spectrometry. The ranges were 5.5 to 49.0 pmol/L in the morning and below 27.5 in the evening 5. Those figures belong to saliva. A serum total testosterone on your printout cannot be checked against them 5.

The same goes for any range you meet online. Outpatients, saliva and a manufacturer's check are different samples, measured different ways. They are not interchangeable.

What to say when two results do not match

Say: "My last result came from a different method. Can we compare the two?"

Ask the laboratory for the method, the lowest value it measures and the source of the range.

Write the old and new results side by side with date, units and method. Bring the sheet. A gap between two printouts may be the machines.

Our column "Who Set Your Testosterone Range, and With What Machine?" is at /female-testosterone-reference-range-2/. A second column on the same range is at /female-testosterone-reference-range/.

Go back to the column headed normal on your printout. It names a range. It never names the women.

If it does not go to plan

  • The portal shows only the number and a range, with no footnotesAsk the laboratory for the full printed report. The method and the source of the range sit in its footer.
  • Your two results use different unitsMultiply nmol/L by 28.8 to reach ng/dL. This converts units only. If the methods also differ, the two numbers still do not line up.
  • The laboratory changed platform between your testsAsk both laboratories to name the method for each date. A change in the number may follow the change in the machine.
  • You are holding a total testosterone result and a free testosterone resultTreat them as two different tests and do not place them against each other. Ask which one each printout reports.

Honestly: what is settled and what is not

Established: in the best-documented case, the female range came indirectly from outpatients on one platform, with a measurement floor of 0.53 nmol/L [[28643405]]. Not established: how a serum range shifts with cycle phase, time of day, age or contraceptives. I found no numbers for that. The figures do not transfer to other platforms, and the papers behind this column do not size the immunoassay difference in women.

What to do with this

  • Compare your number only with a range from the same method, because a printed range belongs to one method and one sample of women
  • Read "below 0.5" as "the instrument cannot see the number" 1
  • Ask which method each result used before guessing whether it reads high or low: immunoassays read higher in one comparison and slightly lower in another 21
  • Write down method, units, floor and range source from every printout, and ask when two results come from different methods

Questions people also ask

Can I compare my testosterone result from one laboratory with another laboratory's range?

Only if both use the same method. A range reflects the method and the women the laboratory sampled 1. Immunoassays read higher than mass spectrometry in one comparison 2 and slightly lower in another 1, and newer versions of the same assay differ from older ones 3. Ask for the method.

What does "<0.5" mean on a testosterone result?

The result is below what the instrument can see. On the platform described here, reliable measurement starts at 0.53 nmol/L 1. The sign says the method could not read the number. It does not say the value is zero or that it is low.

Why do saliva testosterone numbers look so different from my blood result?

Saliva and serum are different samples measured differently. In 223 adult women the morning saliva range was 5.5 to 49.0 pmol/L 5. Those figures cannot be checked against a serum total testosterone printout.

Sources

Every number above comes from one of these. Each carries its PMID, the number it is filed under in the US National Library of Medicine, if you want to look it up yourself.

  1. Reassessment of the Access Testosterone chemiluminescence assay and comparison with LC-MS method.J Clin Lab Anal, 2018 · PMID 28643405
  2. The Simultaneous measurement of serum testosterone and 5α-dihydrotestosterone by gas chromatography-mass spectrometry (GC-MS).Clin Chim Acta, 2018 · PMID 29122541
  3. Multicenter evaluation of a new automated electrochemiluminescence immunoassay for the quantification of testosterone compared to liquid chromatography tandem mass spectrometry.Clin Biochem, 2011 · PMID 20923677
  4. Determination of testosterone in serum by liquid chromatography-tandem mass spectrometry.Scand J Clin Lab Invest, 2008 · PMID 17852804
  5. Salivary testosterone reference intervals for adult women and men.Clin Chem Lab Med, 2026 · PMID 41996543

This article is information, not medical advice, diagnosis or treatment. Reference ranges, doses and protocols differ between people and between laboratories. Talk to a qualified clinician before changing anything about your care.

Max Zimin

I am not a clinician, and that is the point: I had no stake in defending the ranges. I read the primary literature and the laboratory guidance, I find out who chose a number and in what year, and I write it down with the document attached. I also write the books sold on this site. Everything in a column can be checked against its source, because that is the only kind of argument worth making here.